The AeroTrak® 9001 Condensation Particle Counter (CPC) is the only water-based, laminar flow, high flow rate instrument on the market. Using patented technology, the AeroTrak provides 10 nm sensitivity at a 2.83 L/min (0.1 CFM) flow rate. With an ultra-low false count rate, the AeroTrak is ideally suited for the low particle concentrations found in ISO Class 1 and Class 2 environments.
- 10nm sensitivity
- 2.83 l/min (0.1 CFM) sample flow
- Low false-count rate (< 1 particle/ft3) provides accurate measurements at low particle concentrations
- Fast response
- Uses water as working fluid, keeping alcohol and organic compounds out of the clean room
- Data Storage: 250 zones, 999 locations, and 10,000 samples
- USB, USB storage device, and Ethernet data transfer
- External reporting via TrakPro™ Lite Secure Software
- Industry’s loudest audible alarm
- Easy, icon-driven touch screen operation
- Allows storage and recall of specific recipes and zone configurations
- Works with optional high –pressure diffuser for measurements of high pressure gases
- Displays up to three environmental parameters
- 316L stainless steel enclosure
| Size Range | 10nm - 1000nm |
| Particle Channel Sizes | Single channel |
| Counting Efficiency | 50% at 10nm, 100% for particles ≥20nm |
| Concentration Limits | 25,000,000 p/Cu. ft @ 10% coincidence |
| Light Source | Long life laser diode |
| Zero Count | <1 count in 10 minutes |
| Flow Rate | 2.83 L/min (0.1 CFM) |
| Flow Rate Control | Electronic automatic closed loop (patented flow control technology) |
| Sample Tube Extension | Up to 3 m (10 ft) for 10 nm particles |
| Sampling Modes | Manual, automatic, beep; count or concentration |
| Sampling Time | 1 second to 99 hours |
| Sampling Frequency |
1 to 9999 cycles or continuous |
| Exhaust | HEPA filtered |
| Vacuum Source | Internal pump |
| Communication Mode | Modbus TCP over Ethernet or USB |
| Audible Alarm | Built-in; > 85 dBA at 1 meter (adjustable) |
| Status Indicators | Flow, laser and temperatures |
| Alarm Output | Dry contacts, closed when alarm is engaged |
| Alarm Limits | Programmable |
| Display | VGA 14.5 cm (5.7 inch) touch screen display |
| Languages | English, German, French, Spanish, Japanese, Chinese (simplified), Italian |
| Software | Compatible with TrakPro Lite Secure and FMS 5 Software |
| Environmental Sensor Interface | Support TSI Probe Models: 960, 962, 964, 966 (air velocity/temp RH) and Temp/RH probe |
| Unit ID | Configurable IP address |
| Security | 2-level password protection to lock out usage and configuration |
| Data Storage |
250 zones |
| 999 Locations | |
| 10,000 sample records including date, time, particle concentration, flow status and instrument status | |
| Data Transfer | Via USB storage device, connection to TrakPro Lite Secure Software through Ethernet of USB cable, or optional TSI FMS Software |
| Calibration Frequency | Recommended minimum of once per year |
| Dimension (HxWxD) | 52.1 cm x 22.9 cm x 21.8 cm (20.5 in x 9.0 in x 8.6 in) (without water bottle) |
| Weight | 12.1 kg (26.7 lbs) |
| Power | 110 to 240 VAC universal power supply |
| External Surface | 316L Stainless steel |
| Operating Range | 10˚ to 40˚C (50˚ to 104˚F) |
| Storage Range | -40˚ to 70˚C (-40˚ to 158˚F) |
| Standards | N/A |
| Warranty | One year |
| Included Accessories | Quick Start Guide, manual on a CD, power supply, purge filter, 2 inlets (barbed inlet and a 3/16” OD straight tube connector inlet), water bottle, bracket for water bottle and USB cable |
| Optional Accessories | TSI velocity probes, Temp/RH probe, sample tubing |